Comparative Dosimetric Analysis of Tandem-Ovoid and Tandem-Ring Brachytherapy Applicators in Cervical Cancer
DOI:
https://doi.org/10.47981/j.mijst.14(01)2026.582(119-128)Abstract
Brachytherapy plays a vital role in the treatment of locally advanced cervical cancer by delivering high radiation doses directly to the tumor while minimizing exposure to surrounding healthy tissues. This retrospective study compared the dosimetric performance of tandem–ovoid (T&O) and tandem–ring (T&R) intracavitary brachytherapy applicators in 60 cervical cancer patients treated at CMH Cancer Center, Dhaka. Patients were divided into two groups according to applicator type, with 30 patients in each arm. Dosimetric parameters including HR-CTV 90% coverage, rectal dose, and bladder dose were analyzed. The average HR-CTV 90% coverage was 106.91% for the tandem–ovoid applicator and 100.43% for the tandem–ring applicator. The average rectal dose was lower in the tandem–ring applicator (4.93 Gy) compared to the tandem–ovoid applicator (5.57 Gy). However, the bladder dose was slightly higher in the tandem–ring applicator (5.31 Gy) than in the tandem–ovoid applicator (5.20 Gy). The findings indicate that both applicators provide clinically acceptable dose distribution, although tandem–ovoid demonstrated comparatively better target coverage while tandem–ring showed improved rectal dose sparing. Applicator selection should therefore be individualized according to treatment priorities and patient anatomy.
Downloads
References
Aijaz, A. K., Vijay, A., Wani, S. Q., & Haq, M. M. (2022). Dosimetric analysis of intracavitary brachytherapy applicators: A practical study. Radiation Oncology Journal, 40(3), 180–191.
https://doi.org/10.3857/roj.2022.00422
Everett, J. M., Kastan, M. B., & Kirsch, D. G. (2013). Strategies for optimizing the response of cancer and normal tissues to radiation. Nature Reviews Drug Discovery, 12(7), 526–542. https://doi.org/10.1038/nrd4003
Ferlay, J., Soerjomataram, I., Dikshit, R., Eser, S., Mathers, C., Rebelo, M., Parkin, D. M., Forman, D., & Bray, F. (2015). Cancer incidence and mortality worldwide: Sources, methods and major patterns in GLOBOCAN 2012. International Journal of Cancer, 136(5), E359–E386. https://doi.org/10.1002/ijc.29210
Furuse, J., Kinoshita, T., Kawashima, M., Ishii, H., Nagase, M., Konishi, M., Nakagohri, T., Inoue, K., Ogino, T., Ikeda, H., Maru, Y., & Yoshino, M. (2003). Intraoperative and conformal external-beam radiation therapy with protracted 5-fluorouracil infusion in patients with locally advanced pancreatic carcinoma. Cancer, 97(5), 1346–1352. https://doi.org/10.1002/cncr.11165
Gursel, S. B., Serarslan, A., Meydan, A. D., Okumus, N., & Yasayacak, T. (2020). A comparison of tandem ring and tandem ovoid treatment as a curative brachytherapy component for cervical cancer. Journal of Contemporary Brachytherapy, 12(2), 111–117. https://doi.org/10.5114/jcb.2020.94308
Hashemi, F. A., Akbari, E. H., Kalaghchi, B., & Esmati, E. (2013). Concurrent chemoradiation with weekly gemcitabine and cisplatin for locally advanced cervical cancer. Asian Pacific Journal of Cancer Prevention, 14(9), 5385–5389. https://doi.org/10.7314/APJCP.2013.14.9.5385
Jeremic, B., Dubinsky, P., Jeremic, M., Kiladze, I., & Özyiğit, G. (2020). Optimal dose of cisplatin given concurrently with radiotherapy in locally advanced squamous cell head and neck cancer. Turkish Journal of Oncology, 35(2), 237–244. https://doi.org/10.5505/tjo.2020.2189
Khan, M., Hussain, M. A. B., Siddiqui, S. A., & Akram, M. (2022). Five-year survival outcomes with two different high-dose-rate brachytherapy schedules used in the treatment of cervical carcinoma. Therapeutic Radiology and Oncology, 6(2), 35–41. https://doi.org/10.1186/s13014-022-02171-3
Lanciano, R. M., Won, M., Coia, L. R., & Hanks, G. E. (1991). Pretreatment and treatment factors associated with improved outcome in squamous cell carcinoma of the uterine cervix. International Journal of Radiation Oncology, Biology, Physics, 20(4), 667–676.
Nag, S., Erickson, B., Thomadsen, B., Orton, C., Demanes, J. D., & Petereit, D. (2000). The American Brachytherapy Society recommendations for high-dose-rate brachytherapy for carcinoma of the cervix. International Journal of Radiation Oncology, Biology, Physics, 48(1), 201–211. https://doi.org/10.1016/S0360-3016(00)00497-1
Nair, M. T., Cheng, M. C., Barker, A., & Rouse, B. S. (1995). High dose rate brachytherapy technique for carcinoma of the cervix using Nucletron applicators. Medical Dosimetry, 20(3), 201–207. https://doi.org/10.1016/0958-3947(95)00024-T
Pötter, R., Tanderup, K., Schmid, M. P., et al. (2021). MRI-guided adaptive brachytherapy in locally advanced cervical cancer (EMBRACE-I). The Lancet Oncology, 22(4), 538–547. https://doi.org/10.1016/S1470-2045(20)30753-1
Rangarajan, R. (2018). Dosimetric evaluation of image-based brachytherapy using tandem ovoid and tandem ring applicators. Reports of Practical Oncology and Radiotherapy, 23, 57–60.
Sheets, N. C., Goldin, G. H., Meyer, A. M., Wu, Y., Chang, Y., Stürmer, T., Holmes, J. A., Reeve, B. B., Godley, P. A., Carpenter, W. R., & Chen, R. C. (2012). Intensity-modulated radiation therapy, proton therapy, or conformal radiation therapy and morbidity and disease control in localized prostate cancer. JAMA, 307(15), 1611–1620. https://doi.org/10.1001/jama.2012.460
Shrivastava, S., Mahantshetty, U., Engineer, R., Tongaonkar, H., Kulkarni, J., & Dinshaw, K. (2013). Treatment and outcome in cancer cervix patients treated between 1979 and 1994. Journal of Cancer Research and Therapeutics, 9(4), 672–679. https://doi.org/10.4103/0973-1482.126480
Singh, M., Muiriel, C. L., & Johnson, L. (2012). Genetically engineered mouse models: Closing the gap between preclinical data and trial outcomes. Cancer Research, 72(11), 2695–2700. https://doi.org/10.1158/0008-5472.CAN-12-0835
Viswanathan, A. N., Creutzberg, C. L., et al. (2012). International brachytherapy practice patterns: A survey of gynecologic cancer intergroup. International Journal of Radiation Oncology, Biology, Physics, 82(1), 250–255. https://doi.org/10.1016/j.ijrobp.2010.10.030
Wolli, M., Kagan, A., & Olch, A. (1991). Comparison of the ring applicator and the Fletcher applicator for HDR gynecological brachytherapy. Selectron Brachytherapy Journal, 2, 25–27.
Downloads
Published
Issue
Section
License
Copyright (c) 2026 Sharmin Iffat Talukder, Md. Mokhlesur Rahman, Md. Shamim Hassan, Jawata Afnan Mourashi, Md. Mahmudul Hasan Manna, Mohamed Bin Awang

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
This journal provides immediate open access to its content on the principle that making research freely available to the public supports a greater global exchange of knowledge. Users are permitted to read, download, copy, distribute, print, search, or link to the full texts of the articles, provided that appropriate credit is given to the original authors.